HPGe Infrared Detectors
Application
High sensitive HPGe infrared detectors are intended for NIR Fluorescence or Raman spectroscopy and similar applications in spectral region from 850 nm to 1.7 µm.
Features
- Highest sensitivity
- Low noise level
- Frequency range up to 300 Hz
- LN2 cooled electronic input stage (FET and feedback resistor)
- RG 850 window with antireflective coating
- Preamplifier with temperature monitor
- Various types of Dewar vessel are available
COMPLETE SET
- HPGe IR detector in cryostat and LN2 Dewar
- Low noise preamplifier with HV bias supply for detector
ACCESSORIES
specification
Parameter | value |
---|---|
Detector area, mm2 | 20 |
Operating Temp, K | 77 |
Noise Equivalent Power (NEP), W | About 10^-16 |
Responsivity, V/W | 5*10(10) |
Output Impedance, Ohm | 10 |
Output Noise Level, μV(rms) | 60 |
Frequency Bandwidth , Hz | 1-2000 |
Dewar Capacity, l | 7 |
Liquid Nitrogen Holding Time, days | 6 |
Detector ready for work after filling with LN2, h | 2 |
Distance of detector from window, mm | 8-10 |
Entrance collimator diameter, mm | 2 |
DESCRIPTION
HPGe infrared detectors are intended for NIR Fluorescence or Raman spectroscopy and similar applications in spectral region from 850 nm to 1.7 µm.