LB 124 SCINT-D Contamination Monitor with Dose Rate Detector

Screen-Shot-2562-06-02-at-17.24.47

The Contamination Monitor with Dose Rate Detector LB 124 SCINT-D is used to measure radioactive alpha and beta-gamma contaminations on surfaces such as floors, walls, desks, objects, clothing or skin as well as gamma dose rate in ambient dose rate equivalent H*(10).

Features

Highlights:

  • Lightweight, easy to handle and rugged instrument
  • Wide temperature range 

Contamination Detector

  • Simultaneous and separate measurement of alpha and beta-gamma contamination with a ZnS(Ag)-scintillation detector
  • High sensitivity and uniform response
  • No counting gas required 

Dose Rate Detector

  • Detector positioned parallel to the front side of the instrument
  • Energy response to H*(10) Ambient Dose Equivalent
  • Halogen quenched Geiger-Müller tube
  • Dynamic Range: 0.1 μSv/h to 20 mSv/h
  • Energy Range: 50 keV to 1.3 MeV
  • Alarm threshold for dose rate measurement
  • Unit selection: μSv/h
  • Dose accumulation is possible
Instrument
DisplayMonochrome LCD 192 x 64 pixel
Electro-luminescence illumination
Radiation detector ZnS(Ag) scintillator + Geiger-Müller tube
Detection of light Photomultiplier (PMT)
Measurement modes α- and β-γ-measurement, simultaneous
and separate,
additional gamma dose rate H*(10)
ratemeter, scaler-timer-mode, clearance
measurement, survey mode
Entrance window’s dimensions 118 mm × 145 mm
Sensitive area 170 cm²
Entrance window’s material 2 x 3 μm Plastic metallized (0.4 mg/cm²)
Protective Grid 80 % Transmission
External dimensions 240 mm × 140 mm × 110 mm (L x W x H)
Weight 1400 g (with batteries)
Alarm Acoustic with adjustable alarm thresholds
Data memory 1000 measured values with date & time
Interface RS232, headphone connection
Power supply 3 × C size batteries or NiMH batteries,
rechargeable by plug type power supply
or alternatively in the wall mounting
bracket
Max. operating time (without illumination) > 50 h with alkaline batteries 7.8 Ah
> 25 h with NiMH batteries 4.5 Ah
Ambient Conditions
Temperature range -20 °C to +40 °C (operation)
-40 °C to +60 °C (storage)
Relative humidity 0 % to 80 % (no condensation)
External pressure 500 to 1300 hPa (operation)
100 to 1300 hPa (storage/transport)
Protection class IP 53 (according to IEC 60529)
Typical Efficiencies
(according to ISO 7503-1)
Contamination Detector:
241Am 44 % (α-channel)
239Pu 41 % (α-channel)
14C 29 % (β-γ-channel)
36Cl 69 % (β-γ-channel)
60Co 58 % (β-γ-channel)
137Cs 71 % (β-γ-channel)
Background approx. 0.1 cps
approx. 10 cps
(α-channel)
(β-γ-channel)
Gamma sensitivity
at 1 µSv/h (137Cs)
not detectable
< 100 cps
(α-channel)
(β-γ-channel)
Spillover < 20 %
< 2 x 10-5
(α- to β-γ-channel)
(β-γ- to α-channel)
Measuring range
(Dead time < 10 %)
0 to 5.000 cps
0 to 50.000 cps
(α-channel)
(β-γ-channel)
Overrange indication at rates
> 50.000 cps
Suface response
uniformity
max. ± 20 %(α-channel with 241Am,
β-γ-channel with 14C point sources)
Gamma Radiation Detector:
Dose rate range 0.1 µSv/h to 20 mSv/h
Energy range 50 keV to 1.3 MeV
Calibration factor 0.625 µSv/h per cps 137Cs
Intr. background approx. 0.07 cps