The LB 134 Universal Monitor II can be used to measure radioactive alpha and beta-gamma contaminations on surface such as floors, walls, desks, objects, clothing or skin as well as gamma dose rate in ambient dose equivalent H*(10). Therefore a large number of detectors can be attached to the basic unit.

Data communication via RS 485 and USB with F²C protocol including PC programm for remote parameter up- and download, data storage and graphical evaluation. Saving data storage, parameter up- and download and firmware upgrade are possible via USB memory stick.

Features

Highlights of the LB 134 UMo II                                                                                       

  • Lightweight, easy to handle, rugged
  • Connetion to all portable detectors
  • Network capability
  • Wide temperature range
  • PC Software with remote application
  • Practical bag with strap

Dose Rate Detector [built-in]

  • Detector positioned parallel to the front
  • Energy response to H*(10) Ambient Dose Equivalent
  • Halogen quenched Geiger-Müller tube
  • Dynamic range: 100 nSv/h to 20 mSv/h
  • Energy range: 50 keV to 1.3 MeV
  • Alarm threshold for dose rate measurement
  • Accumulated dose and dose rate
    Instrument
    Display Monochrome LCD 192 x 64 pixel
    Electro-luminescence illumination
    Gamma radiation detector Geiger-Müller tube (built-in)
    Measurement modes ratemeter, scaler-timer-, search mode
    External dimensions 160 x 160 x 55 (L x W x D in mm)
    Weight 1400 g (with batteries)
    Data memory 2400 measured values with date & time
    Communication USB, Host for USB stick, RS 485
    Max. operating time > 15 h alkaline batteries 2.6 Ah (4 x AA)
    > 10 h NiMH rechargeable batteries 1.9 Ah
    (internal detector activated)
    Ambient Conditions
    Temperature range – 20 °C to + 40 °C (operation)
    Rel. humidity 0% to 80% (no condensation)
    External pressure 500 to 1300 hPa (operation)
    Protection class IP 53 (according to IEC 60529)
    Sensitivity Gamma Radiation Detector
    Dose rate range 100 nSv/h to 20 mSv/h
    Energy range 50 keV to 1.3 MeV
    Calibration factor 0.625 μSv/h per cps Cs137
    Intr. background approx. 0.07 cps

 