The LB 134 Universal Monitor II can be used to measure radioactive alpha and beta-gamma contaminations on surface such as floors, walls, desks, objects, clothing or skin as well as gamma dose rate in ambient dose equivalent H*(10). Therefore a large number of detectors can be attached to the basic unit.
Data communication via RS 485 and USB with F²C protocol including PC programm for remote parameter up- and download, data storage and graphical evaluation. Saving data storage, parameter up- and download and firmware upgrade are possible via USB memory stick.
Features
Highlights of the LB 134 UMo II
- Lightweight, easy to handle, rugged
- Connetion to all portable detectors
- Network capability
- Wide temperature range
- PC Software with remote application
- Practical bag with strap
Dose Rate Detector [built-in]
- Detector positioned parallel to the front
- Energy response to H*(10) Ambient Dose Equivalent
- Halogen quenched Geiger-Müller tube
- Dynamic range: 100 nSv/h to 20 mSv/h
- Energy range: 50 keV to 1.3 MeV
- Alarm threshold for dose rate measurement
- Accumulated dose and dose rate
Instrument Display Monochrome LCD 192 x 64 pixel
Electro-luminescence illuminationGamma radiation detector Geiger-Müller tube (built-in) Measurement modes ratemeter, scaler-timer-, search mode External dimensions 160 x 160 x 55 (L x W x D in mm) Weight 1400 g (with batteries) Data memory 2400 measured values with date & time Communication USB, Host for USB stick, RS 485 Max. operating time > 15 h alkaline batteries 2.6 Ah (4 x AA)
> 10 h NiMH rechargeable batteries 1.9 Ah
(internal detector activated)Ambient Conditions Temperature range – 20 °C to + 40 °C (operation) Rel. humidity 0% to 80% (no condensation) External pressure 500 to 1300 hPa (operation) Protection class IP 53 (according to IEC 60529) Sensitivity Gamma Radiation Detector Dose rate range 100 nSv/h to 20 mSv/h Energy range 50 keV to 1.3 MeV Calibration factor 0.625 μSv/h per cps Cs137 Intr. background approx. 0.07 cps